Comparison of Heat Generation During Implant Drilling Using Stainless Steel and Ceramic Drills

Research Projects

Organizational Units

Journal Issue

Abstract

Purpose The purpose of this study was to compare the heat generated from implant drilling using stainless steel and ceramic drills. Materials and Methods A total of 40 fresh bovine femoral cortical bone samples were used in this study. A constant drill load of 2.0 kg was applied throughout the drilling procedures via a drilling rig at a speed of 1,500 rpm. Two different implant drill types (stainless steel and ceramic) were evaluated. Heat was measured with type K thermocouple from 3 different depths. Data were subjected to the independent-sample t test and Pearson correlation analysis. The α level was set a priori at 0.05. Results The mean maximum temperatures at the depths of 3 mm, 6 mm, and 9 mm with the stainless steel drill were 32.15°C, 35.94°C, and 37.05°C, respectively, and those with the ceramic drill were 34.49°C, 36.73°C, and 36.52°C, respectively. A statistically significant difference was found at the depth of 3 mm (P = .014) whereas there was no significant difference at the depths of 6 and 9 mm (P > .05) between stainless steel and ceramic drills. Conclusion Within the limitations of the study, although more heat was generated in the superficial part of the drilling cavity with the ceramic drill, heat modifications seemed not to be correlated with the drill type, whether stainless steel or ceramic, in the deep aspect of the cavity. Further clinical studies are required to determine the effect of drill type on heat generation. © 2011 American Association of Oral and Maxillofacial Surgeons.

Description

Keywords

Citation

WoS Q

Q2

Scopus Q

Q2

Source

Journal of Oral and Maxillofacial Surgery

Volume

69

Issue

5

Start Page

1350

End Page

1354

Endorsement

Review

Supplemented By

Referenced By