Publication:
Effect of Multiple Autoclave Cycles on the Surface Roughness of Hyflex CM and Hyflex EDM Files: An Atomic Force Microscopy Study

dc.authorscopusid57193440556
dc.authorscopusid57191441826
dc.authorscopusid56348812200
dc.contributor.authorYılmaz, K.
dc.contributor.authorUslu, G.
dc.contributor.authorOzyurek, T.
dc.date.accessioned2020-06-21T13:06:02Z
dc.date.available2020-06-21T13:06:02Z
dc.date.issued2018
dc.departmentOndokuz Mayıs Üniversitesien_US
dc.department-temp[Yılmaz] Koray, Çorum Oral and Dental Health Centre, Corum, Corum, Turkey; [Uslu] Gulşah, Special Dental Clinic, Canakkale, Turkey; [Ozyurek] Taha, Department of Endodontics, Ondokuz Mayis Üniversitesi, Samsun, Turkeyen_US
dc.description.abstractAim: To compare the effect of autoclave cycles on the surface topography and roughness of HyFlex CM and HyFlex EDM instruments using atomic force microscopy (AFM) analysis. Methodology: Eight new files of each brand were subdivided into four subgroups (n = 2/each subgroup). One group was allocated as the control group and not subjected to autoclave sterilization. The other three groups were subjected to different numbers (1, 5, and 10) of autoclave sterilization cycles. After the cycle instruments were subjected to AFM analysis. Roughness average (Ra) and the root mean square (RMS) values were chosen to investigate the surface features of endodontic files. The data was analyzed using one-way ANOVA and post hoc Tamhane tests at 5% significant level. Results: The lowest Ra and RMS values were observed in the HyFlex EDM files that served as the control and in those subjected to a single cycle of autoclave sterilization (P < 0.05). The highest Ra and RMS values were observed in the HyFlex CM and HyFlex EDM files that were subjected to 10 cycles of autoclave sterilization (P < 0.05). The surface roughness values of the HyFlex CM group showed a significant increase after ten autoclave cycles, whereas those of the HyFlex EDM group exhibited a significant change after five autoclave cycles (P < 0.05). Conclusions: Although the initial surface roughness values of the HyFlex EDM files were lower than those of the HyFlex CM files, the surface roughness values of the EDM files showed a statistically significant increase after 5 cycles of autoclave sterilization. In contrast, the surface roughness values of the HyFlex CM files did not increase until 10 cycles of autoclave sterilization. Clinical relevance: Present study indicated that autoclave sterilization negatively affected the surface roughness of the tested NiTi files. © 2018, Springer-Verlag GmbH Germany, part of Springer Nature.en_US
dc.identifier.doi10.1007/s00784-018-2382-5
dc.identifier.endpage2980en_US
dc.identifier.issn1432-6981
dc.identifier.issn1436-3771
dc.identifier.issue9en_US
dc.identifier.pmid29442187
dc.identifier.scopus2-s2.0-85041925965
dc.identifier.scopusqualityQ1
dc.identifier.startpage2975en_US
dc.identifier.urihttps://doi.org/10.1007/s00784-018-2382-5
dc.identifier.volume22en_US
dc.identifier.wosWOS:000449496900002
dc.identifier.wosqualityQ1
dc.language.isoenen_US
dc.publisherSpringer Verlag service@springer.deen_US
dc.relation.ispartofClinical Oral Investigationsen_US
dc.relation.journalClinical Oral Investigationsen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectAFMen_US
dc.subjectAutoclave Sterilizationen_US
dc.subjectHyflex CMen_US
dc.subjectHyflex EDMen_US
dc.subjectNickel–Titaniumen_US
dc.titleEffect of Multiple Autoclave Cycles on the Surface Roughness of Hyflex CM and Hyflex EDM Files: An Atomic Force Microscopy Studyen_US
dc.typeArticleen_US
dspace.entity.typePublication

Files