Publication: Flow-Injection Spectrophotometric Determination of Copper (II) Ion at Low μg L-1 Levels
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Abstract
A highly selective, sensitive and fast determination of trace amounts of copper (II) ion is described. The method is based on the formation of the coloured Cu(II)-N,N'-o-phenylenebis(salicylaldimine) (PBS) complex, which is then selectively monitored at 415 nm. To increase the interaction between copper(II) and the colour forming reagent and preconcentrate copper(II), a microcolumn containing strong cation-exchange resins was placed between the injection manifold and the spectrophotometer. The linear range of the proposed method was 25-100 mug L-1, and the limit of detection (signal-to-noise ratio = 2) was <1 <mu>g L-1 for 20 muL injection volume of copper (II) ion solution at a sampling rate of >50 h(-1). The relative standard deviation at the level of 25-100 mug L-1 was 0.24% for n=5. The precision and accuracy of the method was checked by analysis of a certified reference material, and the applied to river and sea water samples.
Description
Citation
WoS Q
Scopus Q
Source
Volume
79
Issue
2
Start Page
111
End Page
119