Publication:
Comparison of Cyclic Fatigue Resistance of ProGlider and One G Glide Path Files

Loading...
Thumbnail Image

Date

Journal Title

Journal ISSN

Volume Title

Research Projects

Organizational Units

Journal Issue

Abstract

Introduction The aim of this study was to compare the cyclic fatigue resistances of nickel-titanium rotary glide path files, which are manufactured as single-file systems, under dynamic model. Methods Twenty ProGlider (16.02) and 20 One G (14.03) nickel-titanium glide path files were included in this study. Cyclic fatigue tests were performed with specifically manufactured dynamic cyclic testing device. The device has an artificially prepared canal with 60° curvature angle and 5-mm curvature radius. The canal inner diameter is 1.5 mm, and its curvature center is located at the coronal 5 mm from the apical. The files were randomly divided into 2 groups (n = 20): group 1, ProGlider; group 2, One G. Time to fracture was recorded, and the number of cycles to failure for each instrument was obtained. Statistical analyses were performed by using independent samples t test. The statistically significant level was set at P < .05. Results The cyclic fatigue resistance of ProGlider files was found to be statistically significantly higher than that of the One G files (P < .05). Conclusions Within the limitations of the present study, it was concluded that the cyclic fatigue resistance of ProGlider, which is manufactured as single-file glide path system, is higher than that of the One G file. © 2016 American Association of Endodontists

Description

Citation

WoS Q

Q1

Scopus Q

Q1

Source

Journal of Endodontics

Volume

42

Issue

10

Start Page

1555

End Page

1558

Endorsement

Review

Supplemented By

Referenced By