Publication: Comparison of Cyclic Fatigue Resistance of ProGlider and One G Glide Path Files
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Introduction The aim of this study was to compare the cyclic fatigue resistances of nickel-titanium rotary glide path files, which are manufactured as single-file systems, under dynamic model. Methods Twenty ProGlider (16.02) and 20 One G (14.03) nickel-titanium glide path files were included in this study. Cyclic fatigue tests were performed with specifically manufactured dynamic cyclic testing device. The device has an artificially prepared canal with 60° curvature angle and 5-mm curvature radius. The canal inner diameter is 1.5 mm, and its curvature center is located at the coronal 5 mm from the apical. The files were randomly divided into 2 groups (n = 20): group 1, ProGlider; group 2, One G. Time to fracture was recorded, and the number of cycles to failure for each instrument was obtained. Statistical analyses were performed by using independent samples t test. The statistically significant level was set at P < .05. Results The cyclic fatigue resistance of ProGlider files was found to be statistically significantly higher than that of the One G files (P < .05). Conclusions Within the limitations of the present study, it was concluded that the cyclic fatigue resistance of ProGlider, which is manufactured as single-file glide path system, is higher than that of the One G file. © 2016 American Association of Endodontists
Description
Keywords
Citation
WoS Q
Q1
Scopus Q
Q1
Source
Journal of Endodontics
Volume
42
Issue
10
Start Page
1555
End Page
1558