Publication: A Rare Cause of Pacemaker Failure: Interatrial Block
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Pacemaker and implantable cardioverter defibrillators (ICD) systems are useful in detection and differentiation of many symptomatic and asymptomatic arrhythmias. In this report, we described a rare condition that caused by failure in detection of a clinical tachyarrhythmia by a dual chamber pacemaker that implanted because of intermittent atrioventricular (AV) block and sinus node disease in a 46-year-old patient. In our case, bidirectional interatrial block was demonstrated; and the symptoms associated with high ventricular rate caused by left atrial tachyarrhythmias relieved after AV node ablation. ©2013 Wiley Periodicals, Inc.
Description
Keywords
Citation
WoS Q
Scopus Q
Source
Annals of Noninvasive Electrocardiology
Volume
18
Issue
5
Start Page
479
End Page
483