Publication:
Cyclic Fatigue Resistance of R-Pilot, WaveOne Gold Glider, and ProGlider Glide Path Instruments

dc.authorscopusid56471565700
dc.authorscopusid22934205900
dc.authorscopusid57194336212
dc.authorscopusid8731395200
dc.contributor.authorKeskin, C.
dc.contributor.authorInan, U.
dc.contributor.authorDemiral, M.
dc.contributor.authorKeleş, A.
dc.date.accessioned2020-06-21T13:06:02Z
dc.date.available2020-06-21T13:06:02Z
dc.date.issued2018
dc.departmentOndokuz Mayıs Üniversitesien_US
dc.department-temp[Keskin] Cangül, Department of Endodontics, Ondokuz Mayis Üniversitesi, Samsun, Turkey; [Inan] Uǧur, Department of Endodontics, Ondokuz Mayis Üniversitesi, Samsun, Turkey; [Demiral] Murat, Cekmekoy Oral and Dental Health Center, Istanbul, Istanbul, Turkey; [Keleş] Ali, Department of Endodontics, Ondokuz Mayis Üniversitesi, Samsun, Turkeyen_US
dc.description.abstractObjective: The aim of the present study was to compare the cyclic fatigue resistance of R-Pilot (VDW; Munich, Germany) with ProGlider (Denstply Sirona; Ballaigues, Switzerland) and WaveOne Gold Glider (Denstply Sirona; Ballaigues, Switzerland) glide path instruments. Materials and methods: R-Pilot, ProGlider, and WaveOne Gold Glider instruments were collected (n = 15) and tested in a dynamic cyclic fatigue test device, which has an artificial canal with 60° angle of curvature and a 5-mm radius of curvature. All instruments were operated until fracture occurred, and both time to fracture (TF) and the lengths of the fractured fragments were recorded. Mean and standard deviations of TF and fragment length were calculated for each reciprocating system. TF data and fractured fragment length data were subjected to one-way ANOVA and post-hoc Tukey tests (P < 0.05). Also a Weibull analysis was performed on TF data. Results: The cyclic fatigue resistance values of the WaveOne Gold Glider and R-Pilot were significantly higher than those of the ProGlider (P < 0.05), with no significant difference between them (P > 0.05). Weibull analysis revealed that WaveOne Gold Glider showed the highest predicted TF value for 99% survival rate, which was followed by R-Pilot and ProGlider. Regarding the length of the fractured tips, there were no significant differences among the instruments (P > 0.05). Conclusions: The reciprocating WaveOne Gold Glider and R-Pilot instruments had significantly higher cyclic fatigue resistance than rotary ProGlider instruments. Clinical relevance: This study reported that novel reciprocating glide path instruments exhibited higher cyclic fatigue resistance than rotating glide path instrument. © 2018, Springer-Verlag GmbH Germany, part of Springer Nature.en_US
dc.identifier.doi10.1007/s00784-018-2391-4
dc.identifier.endpage3012en_US
dc.identifier.issn1432-6981
dc.identifier.issn1436-3771
dc.identifier.issue9en_US
dc.identifier.pmid29455371
dc.identifier.scopus2-s2.0-85042072226
dc.identifier.scopusqualityQ1
dc.identifier.startpage3007en_US
dc.identifier.urihttps://doi.org/10.1007/s00784-018-2391-4
dc.identifier.volume22en_US
dc.identifier.wosWOS:000449496900006
dc.identifier.wosqualityQ1
dc.language.isoenen_US
dc.publisherSpringer Verlag service@springer.deen_US
dc.relation.ispartofClinical Oral Investigationsen_US
dc.relation.journalClinical Oral Investigationsen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectDynamic Cyclic Fatigue Test Modelen_US
dc.subjectNickel-Titaniumen_US
dc.subjectPathfile Managementen_US
dc.subjectReciprocationen_US
dc.titleCyclic Fatigue Resistance of R-Pilot, WaveOne Gold Glider, and ProGlider Glide Path Instrumentsen_US
dc.typeArticleen_US
dspace.entity.typePublication

Files