dc.description.abstract | AMAÇ: Hipertansiyonun tanı ve takibinde doğru ölçen bir sfigmomanometre kullanılmalıdır. Bu çalışmanın amaçları hastahanede bulunan kan basıncı ölçüm aletlerinin doğruluğunu değerlendirmek ve kan basıncını ölçen sağlık personelini bu konuda bilgilendirmektir. GEREÇ ve YÖNTEMLER: 119 aneroid ve 7 adet cıvalı aletin dahil edildiği çalışma 2009 yılında yapıldı. Bir anket uygulanarak personelin konu ile ilgili bilgisi ve aletler hakkında bilgi sorgulandı. Aletlerin doğruluğu bir Y bağlantı sistemi kullanılarak araştırıldı. Ayrıca aletlerin mekanik özellikleri gözden geçirildi. BULGULAR: Toplam 56 çalışan 126 aletle ilgili ankete yanıt verdi, 53 çalışan (%95) sfigmomanometrelerin doğru ölçüm yapıp yapmadığının kontrol edilmesi gerektiğini biliyordu. Aletin doğruluğunun nasıl kontrol edileceğini bilen kişi sayısı ise 6 (%11) idi. Aletlerin hepsinin manşet boyutu standarttı. 119 aneroid aletten 49'u (%41) hatalı idi. Hata, aneroid aletlerin sadece 5 tanesinde (%4) 10 mmHg'nın üzerinde idi. Aneroid aletlerin doğruluğu ile kullanım süresi arasında istatistiksel olarak anlamlı bir fark vardı. (p0.01). Kullanılan civalı aletlerin hiçbirinde mekanik bir sorun yoktu. 126 aletin 77'si (%61) doğru ölçüm yapıyordu. SONUÇ: Hastahanelerde kullanılan sfigmomanometrelerin doğruluğu konusunda yapılan çalışma sayısı sınırlıdır. Yapılan az sayıdaki çalışma bizim çalışmamızda olduğu gibi sorunun önemini göstermektedir. Hastahanelerde kullanılan sfigmomanometrelerin düzenli olarak doğrulukları test edilmeli ve personel bu konuda eğitilmelidir. | en_US |
dc.description.abstract | OBJECTIVE: An accurate sphygmomanometer should be used in the diagnosis and follow up of hypertension. The aims of this study were to evaluate accuracy of hospital blood pressure measuring devices and to inform health care staff measuring blood pressure. MATERIAL and METHODS: The study which included 119 aneroid and seven mercury sphygmomanometers was conducted in 2009. The knowledge of the staff and information about devices was determined with a questionnaire. The accuracy of the devices was checked by a Y-connection system. In addition, the mechanical features of the devices were also evaluated. RESULTS: A total of 56 staff responded to the questionnaire about 126 devices and 53 staff (95%) knew that sphygmomanometers should be checked for accuracy. Six (11%) staff knew how the devices should be checked. All cuff sizes were standard. 49 (41%) of the aneroid devices were inaccurate. Inaccuracy was greater than 10 mm Hg in only five (4%) of the devices. There was a statistical relationship between accuracy and usage duration of the devices (p<0.01). None of the mercury spyhgmomanometers had a mechanical problem. Of the 126 devices, 77 (61%) were accurate. CONCLUSION: The number of studies about hospital sphygmomanometers is limited. The limited number of studies emphasize the importance of the subject as in our study. The accuracy of spyhgmomanometers used in hospitals should be tested regularly and staff should be trained about this subject. | en_US |
dc.language.iso | tur | en_US |
dc.rights | info:eu-repo/semantics/openAccess | en_US |
dc.subject | Cerrahi | en_US |
dc.title | Hastahane sfigmomanometrelerinin ölçüm değerleri doğru mu? | en_US |
dc.title.alternative | Are measurement readings of hospital sphygmomanometers accurate? | en_US |
dc.type | other | en_US |
dc.contributor.department | OMÜ | en_US |
dc.identifier.volume | 19 | en_US |
dc.identifier.issue | 2 | en_US |
dc.identifier.startpage | 108 | en_US |
dc.identifier.endpage | 112 | en_US |
dc.relation.journal | Türk Nefroloji Diyaliz ve Transplantasyon Dergisi | en_US |
dc.relation.publicationcategory | Diğer | en_US |