dc.description.abstract | Aim: The aim of this study is to evaluate R-Endo R3 file's resistance to cyclic fatigue in comparison with continuous rotation movement and 'Adaptive Motion' (AM) movement under dynamic model. Materials and Methods: Forty pieces of R-Endo R3 (25.04) files (lot no: 062011) were included in the study. Cyclic fatigue tests were performed with specifically manufactured dynamic cyclic testing device. The device has an artificially prepared canal with 60° curvature angle and 5 mm curvature radius. The canal inner diameter is 1.5 mm and the center of curvature is located 5 mm coronally from the apex. The files were randomly divided into two groups (n = 20) and the following procedures were performed - Group 1: Rotary motion (RM), Group 2: AM. Time to fracture was recorded and the number of cycles to failure for each instrument was obtained. Statistical analyses were performed using Student's t-test. The statistical significant level was set at P < 0.05. Results: AM group had a significantly higher cyclic fatigue resistance compared with RM group (P < 0.001). Conclusion: Within the limitation of the present study, it was found that 'Adaptive Motion' statistically increases the resistance to cyclic fatigue of R-Endo R3 files, manufactured for retreatment. © 2017 Saudi Endodontic Journal Published by Wolters Kluwer - Medknow. | en_US |
dc.language.iso | eng | en_US |
dc.publisher | Medknow Publications | en_US |
dc.relation.isversionof | 10.4103/1658-5984.205123 | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Adaptive motion | en_US |
dc.subject | dynamic cyclic fatigue test | en_US |
dc.subject | endodontics | en_US |
dc.subject | nickel-titanium | en_US |
dc.subject | R-Endo | en_US |
dc.title | Effect of adaptive motion on cyclic fatigue resistance of R-Endo nickel-titanium file | en_US |
dc.type | article | en_US |
dc.contributor.department | OMÜ | en_US |
dc.identifier.volume | 7 | en_US |
dc.identifier.issue | 2 | en_US |
dc.identifier.startpage | 82 | en_US |
dc.identifier.endpage | 86 | en_US |
dc.relation.journal | Saudi Endodontic Journal | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |