Comparison of Cannulated Screw With Tension Band Wiring Versus Compressive Cannulated Locking Bolt and Nut Device (CompresSURE) in Patella Fractures-A Cadaveric Biomechanical Study

Tarih

2012

Yazar

Domby, Brian
Henderson, Eric
Nayak, Aniruddh
Erdogan, Murat
Gutierrez, Sergio
Santoni, Brandon G.
Sagi, H. Claude

Üst veri

Tüm öğe kaydını göster

Özet

Objective: The purpose of this study was to determine if proximity of cannulated lag screws to the articular surface of the patella combined with the tension band technique affects resistance to fracture gap opening, and if an alternative locking nut and bolt device without a tension band behaves in a biomechanically similar fashion. Methods: Thirty-three cadaveric knees were allocated to 3 transverse patella fracture fixation groups: cannulated lag screw placement close to the articular (TBA) or non-articular (TBNA) surface with tension, and placement of a compressive locking nut and bolt device without tension band (CompresSURE) close to the articular surface. Knees were cycled through flexion-extension motion for 1000 cycles during which the fracture gap opening was quantified after the first flexion-extension cycle and after the 1000th cycle using an optoelectronic motion analysis system. Results: After the first range of motion cycle, there was no significant difference in fracture gap opening between the 3 groups on the articular surface (P > 0.600). Total fracture gap displacement after the 1000th cycle was not significantly different between groups (P > 0.408). In general and irrespective of fixation technique, fractures opened in a wedge-like fashion with larger measured gap on the ventral surface relative to the articulating surface. Conclusions: When combined with the tension band, the proximity of cannulated lag screws to the articular surface did not affect resistance to fracture gap opening. Additionally, the stand-alone CompreSURE cannulated locking nut and bolt device without tension band was able to resist fracture gap opening in transverse fractures as effectively as the cannulated screw with tension band technique.

Kaynak

Journal of Orthopaedic Trauma

Cilt

26

Sayı

12

Bağlantı

https://doi.org/10.1097/BOT.0b013e31826f5985
https://hdl.handle.net/20.500.12712/16230

Koleksiyonlar

  • PubMed İndeksli Yayınlar Koleksiyonu [6144]
  • Scopus İndeksli Yayınlar Koleksiyonu [14046]
  • WoS İndeksli Yayınlar Koleksiyonu [12971]



DSpace software copyright © 2002-2015  DuraSpace
İletişim | Geri Bildirim
Theme by 
@mire NV
 

 




| Politika | Rehber | İletişim |

DSpace@Ondokuz Mayıs

by OpenAIRE

Gelişmiş Arama

sherpa/romeo

Göz at

Tüm DSpaceBölümler & KoleksiyonlarTarihe GöreYazara GöreBaşlığa GöreKonuya GöreTüre GöreDile GöreBölüme GöreKategoriye GöreYayıncıya GöreErişim ŞekliKurum Yazarına GöreBu KoleksiyonTarihe GöreYazara GöreBaşlığa GöreKonuya GöreTüre GöreDile GöreBölüme GöreKategoriye GöreYayıncıya GöreErişim ŞekliKurum Yazarına Göre

Hesabım

GirişKayıt

İstatistikler

Google Analitik İstatistiklerini Görüntüle

DSpace software copyright © 2002-2015  DuraSpace
İletişim | Geri Bildirim
Theme by 
@mire NV
 

 


|| Politika || Kütüphane || Ondokuz Mayıs Üniversitesi || OAI-PMH ||

Ondokuz Mayıs Üniversitesi, Samsun, Türkiye
İçerikte herhangi bir hata görürseniz, lütfen bildiriniz:

Creative Commons License
Ondokuz Mayıs Üniversitesi Institutional Repository is licensed under a Creative Commons Attribution-NonCommercial-NoDerivs 4.0 Unported License..

DSpace@Ondokuz Mayıs:


DSpace 6.2

tarafından İdeal DSpace hizmetleri çerçevesinde özelleştirilerek kurulmuştur.