dc.contributor.authorDemircan, Sabri
dc.contributor.authorYuksel, Serkan
dc.contributor.authorGulel, Okan
dc.date.accessioned2020-06-21T14:04:43Z
dc.date.available2020-06-21T14:04:43Z
dc.date.issued2013
dc.identifier.issn1082-720X
dc.identifier.issn1542-474X
dc.identifier.urihttps://doi.org/10.1111/anec.12031
dc.identifier.urihttps://hdl.handle.net/20.500.12712/15701
dc.descriptionYuksel, Serkan/0000-0001-9501-4568en_US
dc.descriptionWOS: 000324522300010en_US
dc.descriptionPubMed: 24047493en_US
dc.description.abstractPacemaker and implantable cardioverter defibrillators (ICD) systems are useful in detection and differentiation of many symptomatic and asymptomatic arrhythmias. In this report, we described a rare condition that caused by failure in detection of a clinical tachyarrhythmia by a dual chamber pacemaker that implanted because of intermittent atrioventricular (AV) block and sinus node disease in a 46-year-old patient. In our case, bidirectional interatrial block was demonstrated; and the symptoms associated with high ventricular rate caused by left atrial tachyarrhythmias relieved after AV node ablation.en_US
dc.language.isoengen_US
dc.publisherWiley-Blackwellen_US
dc.relation.isversionof10.1111/anec.12031en_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectablationen_US
dc.subjectatrial tachyarrhythmiaen_US
dc.subjectinteratrial blocken_US
dc.subjectpacemakeren_US
dc.titleA Rare Cause of Pacemaker Failure: Interatrial Blocken_US
dc.typearticleen_US
dc.contributor.departmentOMÜen_US
dc.identifier.volume18en_US
dc.identifier.issue5en_US
dc.identifier.startpage479en_US
dc.identifier.endpage483en_US
dc.relation.journalAnnals of Noninvasive Electrocardiologyen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US