dc.description.abstract | Pacemaker and implantable cardioverter defibrillators (ICD) systems are useful in detection and differentiation of many symptomatic and asymptomatic arrhythmias. In this report, we described a rare condition that caused by failure in detection of a clinical tachyarrhythmia by a dual chamber pacemaker that implanted because of intermittent atrioventricular (AV) block and sinus node disease in a 46-year-old patient. In our case, bidirectional interatrial block was demonstrated; and the symptoms associated with high ventricular rate caused by left atrial tachyarrhythmias relieved after AV node ablation. | en_US |
dc.language.iso | eng | en_US |
dc.publisher | Wiley-Blackwell | en_US |
dc.relation.isversionof | 10.1111/anec.12031 | en_US |
dc.rights | info:eu-repo/semantics/openAccess | en_US |
dc.subject | ablation | en_US |
dc.subject | atrial tachyarrhythmia | en_US |
dc.subject | interatrial block | en_US |
dc.subject | pacemaker | en_US |
dc.title | A Rare Cause of Pacemaker Failure: Interatrial Block | en_US |
dc.type | article | en_US |
dc.contributor.department | OMÜ | en_US |
dc.identifier.volume | 18 | en_US |
dc.identifier.issue | 5 | en_US |
dc.identifier.startpage | 479 | en_US |
dc.identifier.endpage | 483 | en_US |
dc.relation.journal | Annals of Noninvasive Electrocardiology | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |